Manual Reprocessing Of Gastroscopes
It is important to know the proper manual reprocessing of Pentax gastroscopes to prevent damage to any of its part.
- Wipe insertion tube with a clean, lint-free cloth saturated with detergent solution immediately after examination.
- Alternately flush detergent solution and air through suction channel.
- Disconnect the endoscope from the processor/light source and suction pump.
- Attach the soaking cap on power supply connector; for specialized devices, please use appropriate soaking cap according to IFU.
- Perform leakage test I − the dry test.
- Brush clean all of the channels that may be cleaned with brushes using the appropriate brush under fluid surface, following the order according to the IFU.
- Start brush cleaning with the power supply connector.
- Remove endoscope and any detached components from the cleaning solution.